AquaStorm Clamp-On Flow Meter Configuration and Display Setup

This article provides step-by-step instructions to configure a Clamp-On Flow Meter in the Aquastorm machine software and to set the flow meter display for Gallons Per Minute (GPM) or Liters Per Minute (LPM).


Configure Clamp-On Flow Meter in Machine Software  

  • Open the machine software.

  • Navigate to: Configuration → Modules

  • Select the appropriate module tab where the flow meter is used:

    • Final Rinse

    • Wash Tank

    • Chemical Isolation

    • Chemical Wash (AS200DW)

  • In the Flow Meter dropdown, select Clamp On.

  • Click OK to save the configuration.


Flow Meter Display Setup – Gallons Per Minute (GPM)

Enter Reset / Programming Mode

  1. From the rear of the machine, access the Flow Meter.
  2. Press and hold MODE, then quickly press the UP ARROW 5 times.

  3. Display shows rSEt / no.

  4. Press UP to change to yES, then press MODE..

  5. Display shows dashed lines, followed by Unit G or P-n.

Enter Code Configuration

  1. Press MODE + UP + DOWN simultaneously.

  2. Display shows CodE / no.

  3. Press UP to select yES. Display changes to 0000 / 0000.

  4. Enter the correct code:

7129 / 1400 → Left-to-Right flow

7138 / 1400 → Right-to-Left flow

10. Use UP/DOWN to change digits and MODE to move to the next digit.

11.When all digits are changed, Press and Hold the MODE button until the display shows CodE/= = = and then when dashes disappear CodE/oK.

Analog Output and Scaling Settings

  1. Press and hold MODE until  appears.out 1 / PuLS

  2. Press MODE to step through and verify/set the following:

DisplayRequired Setting
AnLG4-20
Hello0.0
A-Hi15.0
SPd1.0
ttL0.1
FuLLFuLL
made0.1
AvEd1
cut0.5
LedGrn
EcooFF
tEStoFF
PlaceStd
you>>r or L<< (match flow direction)
1/2PiPE
AdJoFF
EndEnd
  1. Final display shows 0.0 or the actual flow rate.

Gallons Per Minute setup is complete.


Flow Meter Display Setup – Liters Per Minute (LPM)

Enter Reset / Programming Mode

  1. From the rear of the machine, access the Flow Meter.
  2. Press and hold MODE, then quickly press the UP ARROW 5 times.

  3. Display shows rSEt / no.

  4. Press UP to change to yES, then press MODE..

  5. Display shows dashed lines, followed by Unit G or P-n.

Enter Code Configuration

  1. Press MODE + UP + DOWN simultaneously.

  2. Display shows CodE / no.

  3. Press UP to select yES. Display changes to 0000 / 0000.

  9. Enter the correct code:

     702C / 0400 → Left-to-Right flow

     703A / 0400 → Right-to-Left flow

10. Use UP/DOWN to change digits and MODE to move to the next digit.

11. When all digits are changed, Press and Hold the MODE button until the display shows CodE/= = = and then when dashes disappear CodE/oK.

Analog Output and Scaling Settings

  1. Press and hold MODE until  appears.out 1 / PuLS

  2. Press MODE to step through and verify/set the following:
DisplayRequired Setting
AnLG4-20
Hello0.0
A-Hi56.8
SPd1.0
ttL0.1
FuLLFuLL
made0.1
AvEd1
cut0.5
LedGrn
EcooFF
tEStoFF
PlaceStd
you>>r or L<< (match flow direction)
1/2PiPE
AdJoFF
EndEnd

14. Final display shows 0.0 or the actual flow rate.

Liters Per Minute setup is complete.