Issue: Weight Scale Communication & Calibration Errors Symptoms

Simultaneous occurrence of two distinct errors depending on the action taken:

  1. "Weight scale not present" — Triggered when attempting to calibrate the weight scale.

  2. "Weight scale communication issue" — Triggered during standard operation (when calibration is not attempted).

Root Cause

The communication protocol for the weight scale was incorrectly modified, resulting in a loss of data transmission with the Benchmark Software.

Resolution

  1. Update Protocol: Verify the scale's communication settings and ensure the correct protocol (xBPI) is configured.

  2. Hardware Replacement: If the communication protocol is already set correctly to xBPI and the errors persist, replace the physical weight scale unit.